  Descriptions :

OLT-55 Intelligent Optical Loss Tester is an all-in-one loss tester integrated with Stabilized Laser Source, Optical Power Meter, Optical Loss Tester, ORL Tester and Length Testing functions. It can perform automatic bidirectional loss test on single fiber with Pass/Fail assessment. With rich functions, flexible configuration and excellent user friendly design, OLT-55 is the ideal optical loss tester to completely evaluate the optical fiber link condition in FTTx, LAN and CATV network.

                  

Main Features:                            

uAll-in-one: SLS + OPM + OLT + ORL + Length test 

uOptical power monitoring (Auto power trace)

uBidirectional loss test on single fiber

uORL test

uPass/Fail assessment

uAutomatic wavelength identification/switch (AutoID)

uRemote reference value setting

uInternal clock & fiber S/N editable

u1000 test records (CSV) storage & management

uData transfer to PC via USB (Driver-free)

uAuto test records save in local unit/remote unit/both units

uUSB power charging

uOver 35 hours continuous operation

uMultilanguage support

uNo warm-up, quick start

uHigh resolution color LCD   

uPocketsize, lightweight and easy-to-use  


 Specifications :

 

Model

OLT-55

Stabilized Laser Source

Wavelength (±20nm)

1310/1550nm

1310/1490/1550nm

1310/1550/1625nm

Spectrum Width

≤5nm

Emitter Type

FP-LD

Output Power

>-4dBm

Power Stability

±0.05dB/15min; ±0.10dB/8hr

Output Mode

CW, 270Hz, 1KHz, 2KHz

Optical Power Meter

Calibrated Wavelength 

850, 1300, 1310, 1490, 1550, 1625nm

Power Range (dBm)

-70 ~ +10 (-60 ~ +10@850nm)

-50 ~ +27

Display Unit

W/mW/μW/nW/pW/dBm/dB(REF)

Detector Type

                           InGaAs

Accuracy

± 5% ± 0.01nW (±0.5 dB@850nm)

± 5% ± 1nW (±0.5 dB@850nm)

Resolution

0.01dB

Auto Wavelength Identification  

Yes (With ShinewayTech specified Stabilized Laser Source)

MOD Identification

270Hz, 1KHz, 2KHz

Optical Loss Test

Link Loss Range

50dB (1550nm, 200km)

Link Loss Accuracy

±0.25dB

Link Loss Test Time

< 2 seconds per wavelength

ORL Test (Optional)

ORL Range

0~60dB (APC connector)

ORL Accuracy

±0.75dB@(0~50dB), ±1.5dB@(50~60dB)

ORL Uncertainty

±0.5dB@20dB

General Specifications

Connector

FC (Interchangeable SC, ST)

Data Storage

1000 records

Data Interface

USB

Backlight Display

Yes

Auto Off

Yes (Auto-off after 5 minutes idle)

Power Supply

Lithium Battery/ AC Adapter

Battery Life

Continuous operation ≥35 hours

Operating Temperature

-20ºC ~ 50ºC

Storage Temperature

-40ºC ~ 70ºC

Relative Humidity

0 ~ 95% (Non-condensing)

Weight

350g

Dimensions (H × W × T)

177×80×44mm 


* Specifications subject to change without notice


 Ordering information:

Standard Package Includes:

card, CE certificate, Certificate of calibration, User manual.